Mindray BeneHeart D3 Defibrillator / Monitor
Mindray BeneHeart D3 defibrillator/monitor with 1-360 J biphasic defibrillation, AED and noninvasive pacing in a 4.7 kg unit built for ambulance and emergency use.
ICEN supplies selected professional medical equipment. Product availability and market authorization vary by country.
A 4-in-1 defibrillator/monitor for ambulance and emergency use
The Mindray BeneHeart D3 combines manual defibrillation, AED, noninvasive pacing and patient monitoring in one unit. Manual output covers 21 selectable energy settings from 1 J to 360 J using a biphasic truncated exponential waveform with automatic impedance compensation, so delivered energy is adjusted to the patient impedance rather than being fixed by the setting alone.
Speed is treated as a specification rather than a claim: the unit powers on in less than two seconds, recovers the ECG trace in less than 2.5 seconds, charges to 200 J in under three seconds and to 360 J in under seven seconds. Synchronized cardioversion begins energy transfer within 60 ms of the QRS peak. The AED shock series is configurable from 100 J to 360 J, and the default configuration meets the 2015 AHA Guidelines.
Monitoring covers 3/5-lead diagnostic ECG with respiration, NIBP, SpO2 and EtCO2, with SpO2 options from Mindray, Nellcor or Masimo. Noninvasive pacing runs from 30 to 210 ppm at 0 to 200 mA in demand or fixed mode. The main unit weighs 4.7 kg with a 15.1 V / 5600 mAh lithium-ion pack rated for roughly 200 shocks at 360 J, six hours of monitoring or 4.5 hours of pacing. ICEN supplies the BeneHeart D3 to hospitals, emergency services and distributors, and confirms configuration and market requirements before quotation.
Applications
- Pre-hospital and ambulance defibrillation
- Emergency department and resuscitation rooms
- In-hospital cardiac arrest response
- AED programmes and transport monitoring
Who is this product for?
- Emergency medical services and ambulance operators
- Hospital emergency and intensive care departments
- Facilities needing one transportable defibrillator/monitor
- Distributors supplying emergency care equipment
Selected Product Highlights
Specifications below are presented from the current ICEN product catalog. Confirm final configuration and technical documents with our team before purchase.
4-in-1: manual defibrillation, AED, noninvasive pacing and monitoring
1-360 J over 21 selectable settings, biphasic truncated exponential waveform
Automatic impedance compensation; patient impedance range 25 to 300 Ohm
Powers on in under 2 s; charges to 200 J in under 3 s and 360 J in under 7 s
AED shock series configurable 100-360 J; default meets 2015 AHA Guidelines
15.1 V / 5600 mAh Li-ion pack: about 200 shocks at 360 J, 6 h monitoring
Mindray BeneHeart D3 Defibrillator / Monitor - key specifications
| Brand | Mindray |
|---|---|
| Model | BeneHeart D3 |
| Category | Defibrillator / Monitor; 4-in-1 integrated design (manual defibrillation, AED, pacing, and monitoring modes) |
| Defibrillation energy | Manual mode output energy: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100, 150, 170, 200, 300, 360 J (21 selectable energy settings); up to 360 J |
| Waveform | Biphasic truncated exponential waveform, with impedance compensation |
| Impedance compensation | Yes - automatic; voltage and energy delivery cycle are adjusted automatically to correspond with impedance. Patient impedance range (external defibrillation): 25 to 300 Ohm. (Note: the official D3 datasheet prints '25 to 300 J'; the equivalent D6 datasheet states '25 to 300 Ohm'.) |
| Charge time | Less than 3 seconds to 200 J; less than 7 seconds to 360 J (with a new, fully charged battery) |
| Synchronized cardioversion | Energy transfer begins within 60 ms of the QRS peak; within 25 ms of the External Sync Pulse |
| Power-on / ECG recovery time | Power on time less than 2 seconds with a new, fully charged battery; ECG recovery time less than 2.5 seconds |
| Display | 7 inch TFT Color LCD; 800 x 480 pixels; max 3 channels of waveforms; wave viewing time max 16 s (ECG) |
| Waveform channels | Display: max 3 channels; Recorder: max 3 channels |
| Monitoring parameters | 3/5-lead ECG (diagnostic quality) with respiration, NIBP, SpO2 and EtCO2. (ECG leads: 3-lead I, II, III; 5-lead I, II, III, aVR, aVL, aVF, V. SpO2 options: Mindray / Nellcor / Masimo.) |
| AED mode | Automatic rhythm analysis with voice and text prompts. AED shock series energy level: 100 to 360 J, configurable; shock series: 1, 2, 3, configurable; default configuration meets 2015 AHA Guidelines; CPR mode with 1-channel ECG monitoring; sensitivity and specificity meets AAMI DF-80 |
| Pacing | Noninvasive pacing. Waveform: monophasic square wave pulse; pulse width 20 ms or 40 ms (+/-5%); refractory period 200 to 300 ms, +/-3% (function of rate); modes: demand or fixed; rate 30 ppm to 210 ppm (+/-1.5%); output 0 mA to 200 mA (+/-5% or 5 mA, whichever is greater); 4:1 pacing function |
| Alarms / clinical decision support | Arrhythmia detection: Yes; Alarms: Yes. AED automatic rhythm analysis with voice and text prompts; Adult/Paediatric mode automatically changes default shock energy, monitoring range and alarm limits |
| Data / event storage | Patient profiles: max 100 patients; Events: up to 1000 events per patient; Waveform storage: up to 24 hours of consecutive ECG waveform; Tabular trends: 72 hours (resolution 1 min); Voice recording: max 180 min in total, max 60 min per patient; Data export to PC via USB flash memory |
| Connectivity | Transmits data through 5G/2.4G WiFi; supports international standard IHE HL7 protocol and DHCP (automatic IP address); USB flash memory data export |
| Battery | Type: 15.1 V, 5600 mAh rechargeable lithium ion battery pack; quantity 1; charge time less than 3 hours to 90% and less than 4 hours to 100% with equipment power off; 5-segment LED capacity indicator; capacity (new, fully charged): monitoring mode 6 hours (5-lead ECG, Resp, SpO2, CO2 and NIBP at 15-minute interval, Wi-Fi disabled), defib mode 200 times at 360 J discharge at intervals of 1 minute without recording, pacing mode 4.5 hours (50 Ohm load, 80 bpm, 60 mA). Battery package weight 0.54 kg |
| Power supply | AC power: line voltage 100 to 240 V~ (+/-10%), current 1.8 to 0.8 A, frequency 50/60 Hz (+/-3 Hz). DC power (through DC-AC inverter): input 12 VDC, power consumption 190 W |
| Protection / ingress rating | Water resistance: IPX4 (without external power); Solids resistance: IP4X |
| Drop resistance | Free fall complies with the requirements of 6.3.4.3, EN1789 (height of fall: 0.75 m); Bump complies with 6.3.4.2, EN1789 (medical devices for use in road ambulances); Shock and vibration complies with 21.102, ISO 9919 |
| Dimensions | 288 mm (w) x 203 mm (d) x 275 mm (h) |
| Weight | Main unit: 4.7 kg; battery package (each): 0.54 kg; external paddle set: 0.86 kg |
| Regulatory / certifications | Safety: complies EN/IEC 60601-1; EMC: complies IEC 60601-1-2; AED sensitivity and specificity: meets AAMI DF-80; transport shock/vibration: ISO 9919; bump/free fall: EN1789 |
Our buying guide sets out how to compare defibrillator/monitor specifications; the models below cover different monitoring depths, display sizes and battery configurations.
Common questions about Mindray BeneHeart D3 Defibrillator / Monitor
What modes does the BeneHeart D3 provide?
It is a 4-in-1 unit: manual defibrillation, AED, noninvasive pacing and patient monitoring.
What defibrillation energy does the D3 deliver?
Manual mode offers 21 selectable energy settings from 1 J to 360 J, using a biphasic truncated exponential waveform with impedance compensation.
How fast does the D3 charge?
Less than 3 seconds to 200 J and less than 7 seconds to 360 J with a new, fully charged battery. The unit powers on in under 2 seconds and recovers the ECG in under 2.5 seconds.
Which parameters can it monitor?
3/5-lead diagnostic ECG with respiration, NIBP, SpO2 and EtCO2. SpO2 options include Mindray, Nellcor and Masimo sensors.
How long does the battery last?
The 15.1 V / 5600 mAh lithium-ion pack is rated for about 200 shocks at 360 J, 6 hours of monitoring with a 5-lead ECG, Resp, SpO2, CO2 and NIBP, or 4.5 hours of pacing.
Can ICEN provide a quotation?
Yes. Send the model, quantity, destination market and required configuration through the Request a Quote button.
Need a quotation or product configuration?